Industry : |
| |
Biotechnology |
| | |
Product : |
| |
Lipid emulsion |
| | |
Test Objective : |
| |
Reduce the particle size to be smaller than the currently used Vortex mixer |
| | |
Processor : |
| |
M-110Y     (Microfluidizer High Shear Processor 장비보기) |
| | |
Chambers : |
| |
- Interaction Chamber (upstream) F20Y (75 um)
- Auxiliary Processing Module (downstream) H30Z (200 um) |
| | |
Number of passes : |
| |
2 |
| | |
Shear per pass : |
| |
6,520,000 sec -1> |
| | |
Process pressure : |
| |
18,000 psi |
| | |
Particle size : |
| |
|
| Passes |
Mean Particle Size (nm) |
|
| Premix |
1,097 |
|
| Vortex |
393 |
|
| 1 Pass |
112 |
|
| 2 Passes |
110 |
|
|
| | |
Summary of results : |
| |
The Microfluidizer materials processor reduced the particle size to a smaller mean than that of the Vortex mixer.